Seronorm™ HbA1c Liquid
Lyophilized, whole blood-based control with HbA1c target values in the physiological and pathological range
Liquid, ready-to-use control
No need for reconstitution. Eliminates potential handling errors.
Behaves similar to a patient sample
Contains more than 95% human whole blood; no preservatives or stabilizers added
Analyt composition | |||
Component | Unit | Level 1 | Level 2 |
HbA1c | mmol/mol Hb | 40 | 70 |
Please note:
Analyte composition and analyte concentration may vary with each lot or instrument. Refer to the package insert of currently available lots for specific analyte and stability claims.
Instruments
Instrument specific values for:
Abbott Architect /Alinity Roche cobas c Siemens Atellica / DCA Vantage
Your device is not listed yet?
Please contact us (
Art. No
-
REF 110 261(3 x 1 ml)Seronorm™ HbA1c Liquid L-1
-
REF 110 262(3 x 1 ml)Seronorm™ HbA1c Liquid L-2
Stability
Unopened:
At ≤ -20 °C are stable until to the expiration date on the label
Once opened:
31 days at 2...8 °C